DUT Power Supplies

DUT Power Supplies

Figure 3-9    DUT power supplies




The rows in the DUT power supplies table correspond to the power supply relays on the ASRU card. Define the configuration as follows:

  • Custom power configurations:
    If the target test system does not have standard power supply configurations, specify all the parameters in each row. Table 3-3 describes the parameters.
  • Standard power configurations:
    If the target test system has standard configurations, simply select the power supply type in the PS Type column. The parameters for the selected type will automatically be filled in.
For information on supported power supply configurations, scroll down to the bottom of the screen. Click the button for a power supply type to see its specifications. The specification are also given in Table 3-6.

 

Power supply configurations must match the hardware installed in the testhead. The power supply connections and types are not configurable via the software. The power supplies are hard wired to resources in the system.

Also note the following:

  • The old power supplies (66xx) cannot be used together with the newer power supplies (N67xx).
  • Existing test programs set up for 6634 power supplies will run without modification on systems with N6746 power supplies.
  • Existing test programs set up for 6621A power supplies will run without modification on systems with N6752 power supplies.
  • Existing test programs set up for 6642 power supplies will run without modification on systems with N6773 power supplies.

Table 3-3    Custom Power Configurations

Column
Description
PS Type
Power supply type (model).1
PS Num
Power supply number. Note that the power supply numbers are different for each module (see Table 3-4). It is recommended that the power supply numbers should match the actual testhead wiring.
Output
The output used on the power supply (Table 3-5).
ASRU Channel
The channel to which the power supply is wired.
Note that each relay can carry a maximum of 5A so it is necessary to wire high current outputs to two ASRU channels using the same power supply number and power supply output.

1   When PS Type ps6752 is selected, the power supply number, output, and ASRU channel are automatically assigned and cannot be changed.

Table 3-4    Module Power Supply Numbers

Module
Power Supply Numbers
Module 3
1, 2, 3, 4, 17, 18
Module 2
5, 6, 7, 8, 19, 20
Module 1
9, 10, 11, 12, 21, 22
Module 0
13, 14, 15, 16, 23, 24

Table 3-5    Power Supply Outputs

Power Supply Model
Output
Voltage @ Current
hp6624
S1
0–7V @ 5A or 7–20V @ 2A
S2
0–7V @ 5A or 7–20V @ 2A
S3
0–20V @ 2A or 2–-50V @ .8A
S4
0–20V @ 2A or 2–-50V @ .8A
hp6621
S1
0–7V @ 10A or 7–20V @ 4A
S2
0–7V @ 10A or 7–20V @ 4A
hp6642
S1
0–20V @ 10A
hp6634
S1
0–100V @ 1A
ps5747
S1
0–50V @ 10 A
ps6751 (Modular)
S1–S4
0–50V @ 0–5A, max 50W
ps6752 (Modular)
S1–S4
0–50V @ 0–10A, max 100W
ps6746 (Modular)
S1–S4
0–100V @ 0–1A, max 100W
ps6773 (Modular)
S1–S2
0–20V @ 0–15A, max 300W

Table 3-6    Standard Configurations

PS Type
Description
hp6621
This supports two configurations:
Select hp6621 in the first row of a module to set up a single module configuration (both outputs wired to a single module).
Select hp6621 in the fifth row to set up a split module configuration (two outputs split across both modules in a bank).
hp6624
This supports two configurations:
Select hp6624 in the first row of a module to set up a single module configuration (all four outputs wired to a single module).
Select hp6624 in the fifth row to set up a split module configuration (two pairs of outputs split across both modules in a bank).
hp6634
Select hp6634 in the fifth or sixth row of a module to assign the single output to ASRU channel 5 or 6 respectively.
hp6642
Select hp6642 in the fifth row of a module to connect the single high current output to ASRU channels 5 and 6.
ps5747
Select ps5747 in the third or fourth row of a module to assign the single output to ASRU channel 3 or 4 respectively.
ps6746
Select ps6746 in the fifth or sixth row of a module to assign the single output to ASRU channel 5 or 6 respectively.
ps6751
This supports two configurations:
  • Select ps6751 in the first row of a module to set up a single module configuration (all four outputs wired to a single module).
  • Select ps6751 in the fifth row to set up a split module configuration (two pairs of outputs split across both modules in a bank).
ps6752
This supports two configurations:
  • Select ps6752 in the first row of a module to set up a single or two-module configuration (two outputs wired to one module or all four outputs wired to two modules).
  • Select ps6752 in the fifth row to set up a split module configuration (four outputs split across four modules).
ps6773
Select ps6773 in the fifth row to set up a split module configuration (two outputs split across both modules in a bank).

Comments

Popular posts from this blog

BT-BASIC commands used

8.10     Some of the most frequently used BT-BASIC commands used are: msi                               Changes default working directory. Mass storage is            Same as “msi” cat                               Catalogs (list)the node names in the specified directory.             get                               Brings the contents of a file into the system workspace.             load                             Same as “get”.             msi$                            Returns the directory pathname of the current working directory.             msi “..”                       Backs up one directory level.             findn                            Locates the next occurrence of the a given sting in the workspace. 8.11           If you wish to invoke the HP Board Graphics Viewer, type board graphics at the BT-BASIC command line and press the “ENTER” key on the keyboard. 8.12           A HP Board Graphics Viewer window should now appear

Perhitungan & Cara Merubah Kumparan Blender Dari 220 V Menjadi 12 V

          Seperti yang telah dijelaskan pada buku “menggulung motor listrik arus bolak-balik, servis peralatan listrik rumah tangga kelompok penggerak dan perbaikan peralatan listrik pertukangan”, bahwa motor penggerak yang digunakan pada perlatan listrik rumah tangga dan pertukangan seperti blender, mixer, bor tembak, gerinda dsb menggunakan jenis motor universal. Motor universal adalah jenis motor listrik yang dapat disuplai dengan sumber listrik arus bolak-balik (AC) dan arus searah (DC). Jadi peralatan-peralatan listrik rumah tangga dan pertukangan tersebut yang biasanya kita suplai dengan sumber listrik AC dari PLN atau Genset sebesar 220 V sebenarnya dapat juga kita suplai dengan sumber listrik DC yang tentunya tegangan juga harus sama yakni 220 V.           Yang menjadi permasalahan bagaimana kalau peralatan listrik rumah tangga atau pertukangan tersebut, sebagai contoh misalkan blender yang ingin digunakan atau dioperasikan pada tempat yang tidak terdapat sumber listrik PLN ata

BT-BASIC command line

8.8       At the BT-BASIC command line type the command  msi  and the directory path, then press the “ENTER” key on the keyboard.  Example:                   msi ‘/hp3070/boards/aspect/main’ 8.9       At the BT-BASIC command line type the command  get ‘testplan’ and press the   ENTER” key on the keyboard.  You should now see the body of the testplan file displayed in the work space of the BT-BASIC window. 8.10     Some of the most frequently used BT-BASIC commands used are:

Autodesk SketchBook Pro 2021 Full Version

BAGAS31 – Sesuai dengan namanya, Autodesk SketchBook Pro 2021 Full Version ini merupakan software digital sketching atau drawing terbaik yang bisa kamu gunakan. Pada versi terbaru kali ini, ada beberapa penambahan fitur yang sangat efektif. Dengan fitur baru tersebut, diharapkan mampu meningkatkan proses sketching maupun drawing kamu. Autodesk SketchBook sendiri sudah bisa kamu dapatkan secara gratis melalui website resminya. Namun untuk kamu yang mau download versi Autodesk Sketchbook Pro, maka bisa langsung download melalui link yang sudah saya sediakan di bawah ini. Download Autodesk SketchBook Pro 2021 Full Version Screenshot: System Requirements: Windows 10 2.5 – 2.9 GHz of Intel or AMD CPU 4 GB of Memory 256 MB Graphics card with OpenGL 2.0 support We recommend that you use a pressure-sensitive tablet and pen for basic features Download: Autodesk SketchBook Pro 2021 Full Version [ FileUp ][ Uptobox ][ UsersDrive ] Jamu Only [ FileUp ][ Uptob

Testhead

4.3         Testhead The testhead is that portion of the tester that supports the PIN, ASRU and Controller cards.   The testhead is divided into two BANKS and each BANK is divided into two MODULES, see figure 2 below.  Bank 1 contains modules 0 and 1, bank 2 contains modules 2 and 3.  The test fixtures are placed on the banks of the tester and locked down for board testing.  The testhead cards interface to the test fixture through the spring loaded pogo pin “nails” at the top edge on each of these card types. 4.4       Support Bay The support bay is a stand-alone cabinet that houses the power supplies for the Unit Under Test.  This bay also houses the test station power distribution unit and test station controller on earlier models. 4.5       Emergency Shutdown Switch The emergency shutdown switch is the large red button located at the lower left corner on the front of the testhead.  It turns off all AC power to the testhead, and is equivalent to turning off the m

Kelebihan dan Kekurangan Saluran Listrik Jenis Saluran Udara dan Saluran Bawah Tanah

Berdasarkan pemasangannya,   saluran distribusi dibagi menjadi dua kategori, yaitu : saluran udara (overhead line) merupakan sistem penyaluran tenaga listrik melalui kawat penghantar yang ditompang pada tiang listrik. Sedangkan saluran bawah tanah (underground cable) merupakan sistem penyaluran tenaga listrik melalui kabel-kabel yang ditanamkan di dalam tanah. 1.    Saluran Bawah Tanah (Underground Lines) Saluran distribusi yang menyalurkan energi listrik melalui kabel yang ditanam didalam tanah. Kategori saluran distribusi seperti ini adalah yang favorite untuk pemasangan di dalam kota, karena berada didalam tanah, maka tidak mengganggu keindahan kota dan juga tidak mudah terjadi gangguan akibat kondisi cuaca atau kondisi alam. Namun juga memilik kekurangan, yaitu mahalnya biaya investasi dan sulitnya menentukan titik gangguan dan perbaikannya. Kedua cara penyaluran memiliki keuntungan dan kerugian masing-masing. Keuntungan yang dapat diperoleh dari suatu jaringan bawah tanah adalah