Calibration > System Calibration > Required Equipment and Tools

  Required Equipment and Tools Table 1    Required Equipment and Tools Description Part Number Digital multimeter (DMM) 34401A, 3455A, 3456A, or 3458A Frequency counter 5335A or 53131A NOTE:  Use only the specific models of DMM and counter listed above; other models will not work with the calibration software. The instruments must be calibrated and traceable to a national standards laboratory (e.g., NIST in the U.S.).   Digital hygrometer or other instrument to measure temperature and humidity   4-meter GPIB cable 10833C 1-meter GPIB cable 10833A Two power cords   Power strip extension   Two 1-meter dual banana (M)-to-dual banana (M) coaxial cables   1-meter BNC(M)-to-BNC(M) coaxial cable   ASRU Card calibration jumpers (one per module) 03066-61628 Pin verification fixture For 2- or 4-module testhead (Provides full coverage for ASRU-N card; backward compatible with E4033A) E4033C For 1-module testhead (Provides full coverage for ASRU-N card; backward compatible with E3771A) E3771C For

Calibrate the ASRU Cards

  Calibrate the ASRU Cards 1    Log on as  calibrate . 2    Press  Testhead Functs  (F3) and then  Testhead Power On  (F1) to power-up and boot the testhead.   The testhead requires a 30-minute warm-up before calibration. 3    Place the pin verification fixture on either bank of the testhead that contains a module. 4    Press  Testhead Functs  (F3) and then  Fixture Lock  (F5) to lock the fixture onto the testhead. 5    After the fixture is installed on the testhead, execute test 4395 from the Testhead Test Number Entry screen. If you do not specify a module number, the test will repeat for all modules in the  config  file. Follow the instructions displayed on the screen. When you see the message: Do you want to continue? (yes or no): Type  yes  (all lowercase), not  Y ,  Yes  or  YES . a    When instructed, enter the model number of the multimeter and set the GPIB address of the multimeter. Table 5    Set GPIB address of multimeter Multimeter model Steps 3455A and 3456A DMMs Use the G

Calibration > ASRU Calibration > ASRU Calibration Process > Set up the Equipment

  Set up the Equipment 1    Remove any test fixture or dust covers that may be on the testhead. If an EFS Board Handler is on the testhead (Mux systems only), remove it. 2    The testhead’s power must be OFF (testhead not booted), but the PDU must be switched on. 3    Rotate the testhead to the service position. 4    Place the DMM near or on top of the testhead. Plug it into a power outlet and turn it on to warm up.   The DMM requires adequate warm-up before it is used for calibration: – one hour for the 34401A – two hours for the 3455A – 30 minutes for the 3456A – four hours for the 3458A 5    Open the module doors, install ASRU Card calibration jumpers on all ASRU Cards (in slot 1 of each module) in the testhead, and close the module doors.

Relays by card type

  Relays Table 16    Relays Part Description 03066-69903 SMT relay socket kit. Includes: 0490-1688 SMT relays x 6 03066-60400    sockets x 6 0490-1814 pack of 12 0490-1804 relays Table 17    Relays by card type Card Type Relay Function/Application Relay Part Number Relay Type 1 System Card MPU Enable 0490-2006 TH Fixture Pull-down, Aux port (vacuum control) 0490-1828 SMT Module Control Card GP Relay 0490-1828 SMT Misc. (ControlXT/XTP, 5V) 0490-1839 or 0490-2788 TH ASRU Revision C Card DUT Power Supply 1–4 0490-2006 TH DUT Power Supply 5–6 (ASRU-C only) 0490-1727 TH DUT sense, MOA, sub-MUX 0490-1804 2 TH Trapped Charge Relay (Discharge K799) 0490-1828 SMT Interconnect MUX, sub-MUX, probe, MOA, source, detector, functional. ports, interconnect MUX 0490-1688 SMT ASRU Card, Rev D DUT Power Supply 1-4 0490-2006 TH DUT Power Supply 5-6 (ASRU-C only) 0490-1727 TH Trapped Charge Relay (Discharge K799) 0490-1828 SMT DUT sense, MOA, sub-MUX,Interconnect MUX, sub-MUX, probe, MOA, source, detector

ASRU-N Card Parts

  Table 6    ASRU-N Card Parts Part Description N1807-66500 ASRU-N Card, new N1807-69500 ASRU-N Card, exchange 2110-0699 5A, 125V fuse 2110-0665 1A, 125V fuse 0490-2006 DUT power supply relay (ch 1–4, 4 ea.) 0490-1727 DUT power supply relay (ch 5–6, 2 ea.) Table 7    ASRU-C Card Parts   Part Description 03066-69532 ASRU-C Card, exchange 03066-66532 ASRU-C Card, new 2110-0699 5A, 125V fuse 2110-0665 1A, 125V fuse 0490-2006 DUT power supply relay (ch 1–4, 4 ea.) 0490-1727 DUT power supply relay (ch 5–6, 2 ea.) 0490-1828 trapped charge relay (5V SMT relay) 0490-1688 Interconnect MUX, sub-MUX relays 0490-1804 relays (SIP relay) Table 8    Parts for ASRU Card, Rev D Part Description 03066-69532 exchange 03066-66532 new 2110-0699 5A, 125V fuse 2110-0665 1A, 125V fuse 0490-2006 DUT power supply relay (ch 1-4, 4 ea.) 0490-1828 trapped charge relay (5V SMT relay) 0490-1839 DUT sense, MOA, sub-MUX, Interconnect MUX, sub-MUX, probe, MOA, source, detector, functional ports, interconnect MUX

Utility Card

  Table 9    Utility Card Part Description N1807-66502 Utility Card, new N1807-69502 Utility Card, exchange 0490-1772 MUX relays 1258-0290 Removeable jumper 0490-2006 DUT power supply relay 2110-0932 5A fuse Table 10    Hybrid-144 Non-multiplexed Pin Card Part Description N1140-69500 Hybrid-144 Non-multiplexed Pin Card, exchange N1140-66500 Hybrid-144 Non-multiplexed Pin Card, new 0490-2063 relay, 300-ohm 2110-0946 10A, 125V fuse 2110-0955 2A, 125V S fuse Table 11    HybridPlus Card  1 Part Description E4000-69552 HybridPlus Standard DD Pin Card, exchange E4000-66552 HybridPlus Standard DD Pin Card, new E4000-69553 HybridPlus Advanced DD Pin Card, exchange E4000-66553 HybridPlus Advanced DD Pin Card, new E4000-69554 HybridPlus High Accuracy DD Pin Card, exchange E4000-66554 HybridPlusHigh Accuracy DD Pin Card, new E4000-69555 HybridPlus Pay-Per-Use DD Pin Card, exchange E4000-66555 HybridPlus Pay-Per-Use DD Pin Card, new 2110-0946 10A fuse subminiature 0490-1839 or 0490-2788 5V MUX rel