Calibrate the ASRU Cards
1 Log on as
calibrate
.2 Press Testhead Functs (F3) and then Testhead Power On (F1) to power-up and boot the testhead.
3 Place the pin verification fixture on either bank of the testhead that contains a module.
4 Press Testhead Functs (F3) and then Fixture Lock (F5) to lock the fixture onto the testhead.
5 After the fixture is installed on the testhead, execute test 4395 from the Testhead Test Number Entry screen. If you do not specify a module number, the test will repeat for all modules in the
config
file.Follow the instructions displayed on the screen. When you see the message:
Do you want to continue? (yes or no):
Type
yes
(all lowercase), not Y
, Yes
or YES
.a When instructed, enter the model number of the multimeter and set the GPIB address of the multimeter.
b When instructed, connect the GPIB cable to the system’s GPIB (e.g., the one that goes to the DUT power supplies; there may or may not be a support bay).
c Connect the Input and Sense ports correctly as shown in Figure 2.
d Ensure that the Front/Rear switch is in the Front position.
e After making the necessary connections and answering the questions, test 4395 will complete the calibration.
6 When one module calibration is completed, execute test 4395 on the remaining modules. Move the pin verification fixture to the other bank as appropriate.
7 After all modules are calibrated, execute test 4396 to verify that the calibrations were recorded on today’s date.
8 Press Testhead Functs (F3) and then Fixture Unlock (F6) to unlock the fixture.
Figure 2 Calibration connections